OPEN VENTILATOR - Specifications

Requirements (MVP)

Numeric specifications

  • Inhalation speed 0.7 to 3 s
  • Exhalation speed 1 to 6 s
  • Resulting BPM from 10 to 35 cycles a minute
  • PEEP Valve control from 1cm/H2O to 30cm/H2O
  • Overpressure Valve (adjustable from 1cm/H2O to 50cm/H2O)
  • Mechanical Breath Volume Adjustment up to at least 1 liter/breath

Further Spartan specifications and Mark II prospect

Specification Spartan model Mark II
Volume control Adjustable on crank-wheel Adjustable on crank-wheel
Ventilation frequency Indirectly by setting speeds Electronic native control
Inhale Speed/flow 12 position Switch Electronicly controlled
Exhale Speed/flow 12 position Switch Electronicly controlled
Inhale/Exhale ratio Indirectly by setting speeds Electronic native control
Maximum pressure Adjustable by water column Adjustable by water column
PEEP Pressure regulator Adjustable by water column Adjustable by water column
PEEP Pressure indicator Optional water column indicator Digital UI or WaterColumn
Inhale Pressure Sensor Not Available Electronic sensor
Exhale Pressure Sensor Not Available Electronic sensor
Flow Sensor Not Available Electronic sensor if available
Inlet air filtration HEPA / Coffee Paper Filter HEPA
Air humidifier HME Filter (recommended) HME Filter (recommended)
Exhaust filtration UV-C Filter + Heat Chamber UV-C Filter + Heat Chamber
Oximeter Not available Electronic sensor if available
Heartbeat sensor Not available Electronic Sensor if available
Assist Control No Probably
Cleanable inside Yes Yes

Success Criteria

Validated Prototype Thanks to SAMTRONICS This team was super professional and lent us one of the pulmonary simulators that they have available at the company.

Status: The design works. It needs to be built by more people.

For more results see here: test results