OPEN VENTILATOR - Specifications
- Inhalation speed 0.7 to 3 s
- Exhalation speed 1 to 6 s
- Resulting BPM from 10 to 35 cycles a minute
- PEEP Valve control from 1cm/H2O to 30cm/H2O
- Overpressure Valve (adjustable from 1cm/H2O to 50cm/H2O)
- Mechanical Breath Volume Adjustment up to at least 1 liter/breath
Specification | Spartan model | Mark II |
---|---|---|
Volume control | Adjustable on crank-wheel | Adjustable on crank-wheel |
Ventilation frequency | Indirectly by setting speeds | Electronic native control |
Inhale Speed/flow | 12 position Switch | Electronicly controlled |
Exhale Speed/flow | 12 position Switch | Electronicly controlled |
Inhale/Exhale ratio | Indirectly by setting speeds | Electronic native control |
Maximum pressure | Adjustable by water column | Adjustable by water column |
PEEP Pressure regulator | Adjustable by water column | Adjustable by water column |
PEEP Pressure indicator | Optional water column indicator | Digital UI or WaterColumn |
Inhale Pressure Sensor | Not Available | Electronic sensor |
Exhale Pressure Sensor | Not Available | Electronic sensor |
Flow Sensor | Not Available | Electronic sensor if available |
Inlet air filtration | HEPA / Coffee Paper Filter | HEPA |
Air humidifier | HME Filter (recommended) | HME Filter (recommended) |
Exhaust filtration | UV-C Filter + Heat Chamber | UV-C Filter + Heat Chamber |
Oximeter | Not available | Electronic sensor if available |
Heartbeat sensor | Not available | Electronic Sensor if available |
Assist Control | No | Probably |
Cleanable inside | Yes | Yes |
Validated Prototype Thanks to SAMTRONICS This team was super professional and lent us one of the pulmonary simulators that they have available at the company.
Status: The design works. It needs to be built by more people.
For more results see here: test results